• PH Sensor

PH Sensor

Short Description:

The new-generation PHTRSJ soil pH sensor solves the shortcomings of traditional soil pH that requires professional display instruments, tedious calibration, difficult integration, high power consumption, high price, and difficult to carry.


Product Detail

Product Tags

Product Instruction

The new-generation PHTRSJ soil pH sensor solves the shortcomings of traditional soil pH that requires professional display instruments, tedious calibration, difficult integration, high power consumption, high price, and difficult to carry.

New soil pH sensor, realizing online real-time monitoring of soil pH.
It adopts the most advanced solid dielectric and large-area polytetrafluoroethylene liquid junction, which is not easy to block and maintenance-free.
High integration, small size, low power consumption, easy to carry.
Realize low cost, low price and high performance.
High integration, long life, convenience and high reliability.
Simple operation.
Support secondary development.
The electrode uses high-quality low-noise cable, which can make the signal output length up to 20 meters without interference.

This product can be widely used in the fields of agricultural irrigation, flower gardening, grassland pasture, rapid soil testing, plant cultivation, scientific experiments and so on.

Technique Parameter

Measuring range 0-14pH
Accuracy ± 0.1pH
Resolution 0.01pH
Response time <10 seconds (in water)
Power supply mode DC 12V 
DC 24V 
Other
Output form Voltage: 0~5V 
Current: 4 ~ 20mA
RS232
RS485
Other
Instrument Line length Standard: 5 meters
Other  
Working environment Temperature 0 ~ 80 ℃
Humidity: 0 ~ 95% RH
Power consumption 0.2W
Housing material waterproof plastic shell
Transmitter size 98 * 66 * 49mm

Calculation Formula

Voltage type (0 ~ 5V output):
D = V / 5 × 14
(D is the measured pH value, 0.00pH≤D≤14.00pH, V is the output voltage (V))

Current type (4 ~ 20mA output):
D = (I-4) / 16 × 14
(D is the measured pH value, 0.00pH≤D≤14.00pH, I is the output current (mA))

Wiring Method          

(1) If equipped with a weather station produced by our company, directly connect the sensor to the corresponding interface on the weather station using a sensor line.
(2) If the transmitter is purchased separately, the cable sequence of the transmitter is:

Line color

Output signal

Voltage type

Current type

Communication

type

Red

Power +

Power +

Power +

Black (green)

Power ground

Power ground

Power ground

Yellow

Voltage signal

Current signal

A+/TX

Blue

 

 

B-/RX

Wiring Method          

PH Sensor1

MODBUS-RTU Protocol

1. The serial format
Data bits 8 bits
Stop bit 1 or 2
Check Digit None
Baud rate 9600 Communication interval is at least 1000ms
2. Communication format
[1] Write device address
Send: 00 10 Adress CRC (5 bytes)
Returns: 00 10 CRC (4 bytes)
Note: 1. The address bit of the read and write address command must be 00.
2. Adress is 1 byte and the range is 0-255.
Example: Send 00 10 01 BD C0
Returns 00 10 00 7C
[2] Read device address
Send: 00 20 CRC (4 bytes)
Returns: 00 20 Adress CRC (5 bytes)
Explanation: Adress is 1 byte, the range is 0-255
For example: Send 00 20 00 68
Returns 00 20 01 A9 C0
[3] Read real-time data
Send: Adress 03 00 00 00 01 XX XX
Note: as shown below:

Code

Function definition

Note

Adress

Station number (address)

 

03

Function code

 

00  00

Initial address

 

00  01

Read points

 

XX XX

CRC Check code, front low later high

 

Returns: Adress 03 02 XX XX XX XX

Code

Function definition

Note

Adress

Station number (address)

 

03

Function code

 

02

Read unit byte

 

XX  XX

Data (high before, low after)

Hex

XX  XX

CRCCheck code

 

To calculate the CRC code:
1. The preset 16-bit register is FFFF in hexadecimal (that is, all are 1). Call this register the CRC register.
2. XOR the first 8-bit data with the lower bit of the 16-bit CRC register and put the result in the CRC register.
3. Shift the contents of the register to the right by one bit (toward the low bit), fill the highest bit with 0, and check the lowest bit.
4. If the least significant bit is 0: repeat step 3 (shift again), if the least significant bit is 1: the CRC register is XORed with the polynomial A001 (1010 0000 0000 0001).
5. Repeat steps 3 and 4 until 8 times to the right, so that the entire 8-bit data has been processed.
6. Repeat steps 2 to 5 for the next 8-bit data processing.
7. The CRC register finally obtained is the CRC code.
8. When the CRC result is put into the information frame, the high and low bits are exchanged, and the low bit is first.

RS485 circuit  

PH Sensor2

Instructions for use  

1. When the sensor leaves the factory, the probe is provided with a transparent protective cover, and the built-in protective liquid protects the probe. When using, please remove the protective cover, fix the filter tank and the sensor, and then use the attached cable tie to wrap the filter in the filter tank. To prevent direct contact between the soil and the probe and damage the probe. In actual use, please ensure that the filter trough and the filter are firmly connected. Do not remove the filter trough and the filter. Insert the probe directly into the soil to avoid making the probe damage and unrepairable.
2. Insert the probe part vertically into the soil. The depth of the probe must be at least covered by the filter. Under normal circumstances, the pH in the air is between 6.2 and 7.8.
3. After burying the sensor, pour a certain amount of water around the soil to be measured, wait for a few minutes, and wait for the water to soak into the probe, then you can read the data on the instrument. Under normal circumstances, the soil is neutral and the pH is between Around 7, the actual pH value of the soil in different places will be different, it should be determined according to the actual situation.
4. The user can use the attached 3 pH reagents and configure according to the configuration method to check whether the product performance is normal.

Precautions

1. In the pipeline in order to ensure the correct electrode measured pH value should be avoided during the measurement of air bubbles caused by inaccurate data;
2. Please check whether the packaging is intact and check whether the product model is consistent with the selection;
3. Do not connect with power on, and then power on after checking the wiring.
4. Do not arbitrarily change the components or wires that have been soldered when the product leaves the factory.
5. The sensor is a precision device. Please do not disassemble it by yourself or touch the surface of the sensor with sharp objects or corrosive liquids during use to avoid damaging the product.
6. Please keep the verification certificate and certificate of conformity, and return it with the product when repairing.

Troubleshooting

1. For analog output, the display indicates that the value is 0 or is out of range. The collector may not be able to obtain the information correctly due to wiring problems. Please check whether the wiring is correct and firm, and the power voltage is normal;
2. If it is not the above reasons, please contact the manufacturer.

Maintenance     

1. The input end of the instrument (measuring electrode socket) must be kept dry and clean to prevent dust and water vapor from entering.
2. Avoid long-term immersion of electrodes in protein solution and acid fluoride solution, and avoid contact with silicone oil.
3. After long-term use of the electrode, if the slope is slightly reduced, the lower end of the electrode can be immersed in a 4% HF solution (hydrofluoric acid) for 3 to 5 seconds, then washed with distilled water and then immersed in 0.1mol / L hydrochloric acid Refresh the electrode.
4. In order to make the measurement more accurate, the electrode must be calibrated frequently and washed with distilled water.
5. The transmitter should be placed in a dry environment or a control box to avoid meter leakage or measurement error caused by water droplets splashing or being wet.


  • Previous:
  • Next:

  • Write your message here and send it to us

    Related products

    • Soil temperature and humidity sensor soil transmitter

      Soil temperature and humidity sensor soil trans...

      Technique Parameter Measurement range soil moisture 0 ~ 100% soil temperature -20 ~ 50 ℃ Soil wet resolution 0.1% Temperature resolution 0.1 ℃ Soil wet accuracy ± 3% Temperature accuracy ± 0.5 ℃ Power supply mode DC 5V  DC 12V  DC 24V  Other Output form Current: 4~20mA  Voltage: 0~2.5V  Voltage: 0~5V     RS232 RS485 TTL Level: (frequency; Pulse width) Other Load ...

    • Bus transmitter Instructions

      Bus transmitter Instructions

      485 Overview 485 is a kind of serial bus which is widely used in industrial communication. 485 communication only needs two wires (line A, line B), long distance transmission is recommended to use shielded twisted pair. Theoretically, the maximum transmission distance of 485 is 4000 feet and the maximum transmission rate is 10Mb/s. The length of the balanced twisted pair is inversely proportional to t...

    • Portable gas sampling pump

      Portable gas sampling pump

      Product Parameters ● Display: Large screen dot matrix liquid crystal display ● Resolution: 128*64 ● Language: English and Chinese ● Shell materials: ABS ● Working principle: Diaphragm self-priming ● Flow: 500mL/min ● Pressure: -60kPa ● Noise: <32dB ● working voltage: 3.7V ● Battery capacity: 2500mAh Li battery ● Stand-by time: 30hours(keep pumping open) ● Charging Voltage: DC5V ● Charging Time: 3~5...

    • Portable Handheld Weather Station

      Portable Handheld Weather Station

      Features  ◆ 128 * 64 large-screen LCD displays temperature, humidity, wind speed, average wind speed, maximum wind speed, wind direction, and air pressure value; ◆ Large-capacity data storage, can store up to 40960 weather data (data recording interval can be set between 1 ~ 240 minutes); ◆ Universal USB communication interface for easy data download; ◆ Only need 3 AA batteries: low power consumption ...

    • Integrated tipping bucket rainfall monitoring station Automatic rainfall station

      Integrated tipping bucket rainfall monitoring s...

      Features  ◆ It can automatically collect, record, charge, work independently, and do not need to be on duty; ◆ Power supply: using solar energy + battery: the service life is more than 5 years, and the continuous rainy working time is more than 30 days, and the battery can be fully charged for 7 consecutive sunny days; ◆ Rainfall monitoring station is a product with data collection, storage and transm...

    • Externally attached ultrasonic level meter

      Externally attached ultrasonic level meter

      Introduction External ultrasonic level meter measurement will be placed in the ultrasonic transducer measured container wall directly below (bottom), without opening, easy to install, does not affect the site production. Transducer external, truly non-contact measurement, suitable for a variety of closed containers toxic, volatile, flammable, explosive, strong pressure, strong corrosive and other liqu...