• WGZ-500B, 2B, 3B, 4000B Portable Turbidity Meter

WGZ-500B, 2B, 3B, 4000B Portable Turbidity Meter

Short Description:

Portable, microcomputer, powerful, automatic calibration, can be connected to a printer.

It is used to measure the degree of scattering of light generated by insoluble particulate matter suspended in water or transparent liquid, and to quantitatively characterize the content of these suspended particulate matter. It can be widely used in turbidity measurement in power plants, purified water plants, water plants, domestic sewage treatment plants, beverage plants, environmental protection departments, industrial water, wine and pharmaceutical industries, epidemic prevention departments, hospitals and other departments.


Product Detail

Product Tags

Features 

Portable, AC and DC power supply, with low voltage indication and automatic shutdown function. The serial RS232 communication interface can be connected with a micro printer.
Microcomputer low-power configuration, touch keyboard, LCD screen with backlight, can display date, time, measurement value and measurement unit at the same time.
Measuring range can be manually selected or automatically switched. The calibration standard value can be set arbitrarily by programming, and 1-7 points can be quickly and arbitrarily selected for automatic calibration.
Equipped with average measurement mode and data query mode, with non-linear processing and smoothing functions of measurement data, and equipped with self-diagnosis information prompts.
Built-in clock memory storage system, real-time storage of measurement and correction data, long-term storage and recall of the latest 20 sets of measurement data.
Multiple measurement modes, preset with NTU, FTU, EBC, degrees (Unit), ppm, mg/L, % and other units.
Equipped with a chromaticity compensation system, which can effectively avoid the interference caused by the color of the sample, and can correctly reflect the concept of turbidity.
100,000 hours long-life high-intensity light source, maintenance-free long-term use, in line with ISO turbidity measurement standards.

Technical indicators 

Product number

WGZ-500B(The original WGZ-2AB type)

WGZ-2B

WGZ-3B

WGZ-4000B

Determination principle

90° scattered light

Minimum indication (NTU)

0.01

0.001

0.01

0.001

Measuring range (NTU)

050

0500

010

0100

0500

010

0100

01000

010

0100

01000

04000

Indication error

±6%(±2F.S

Repeatability

≤0.5%

Zero drift

±0.5F.S

Power supply

DC 1.5V×5 AA alkaline dry batteries AC 220V/50Hz/DC7.5V/0.2A  power adapter

Features Microcomputer configuration, with average measurement mode, year, month, date and time display, with data storage and query functions, automatic range switching, automatic zero adjustment and automatic calibration from 1 to 5 points, equipped with RS232 data communication interface, can be connected to external micro printer.

  • Previous:
  • Next:

  • Write your message here and send it to us

    Related products

    • Portable Multiparameter Transmitter

      Portable Multiparameter Transmitter

      Product advantages 1. One machine is multi-purpose, which can be expanded to use various types of sensors; 2. Plug and play, automatically identify electrodes and parameters, and automatically switch the operation interface; 3. The measurement is accurate, the digital signal replaces the analog signal, and there is no interference; 4. Comfortable operation and ergonomic design; 5. Clear interface and ...

    • Single-point Wall-mounted Gas Alarm (Chlorine)

      Single-point Wall-mounted Gas Alarm (Chlorine)

      Technical parameter ● Sensor: catalytic combustion ● Responding time: ≤40s (conventional type) ● Work pattern: continuous operation, high and low alarm point(could be set) ● Analog interface: 4-20mA signal output[option] ● Digital interface: RS485-bus interface [option] ● Display mode: Graphic LCD ● Alarming mode: Audible alarm -- above 90dB; Light alarm -- High intensity strobes ● Output control: rel...

    • Indoor temperature and humidity sensor

      Indoor temperature and humidity sensor

      1、Features ◆The real-time temperature and humidity data on site can be displayed after power-on, without the help of computers and other equipment; ◆High-definition LCD display, the data is clearly visible; ◆Automatically switch real-time temperature and humidity data without manual switching and adjustment; ◆The system is stable, there are few external interference factors, and the data is accurate; ◆Small size, easy to carry and fix. 2、Scope of application Widely used in supermarkets, fac...

    • Portable combustible gas leak detector

      Portable combustible gas leak detector

      Product Parameters ● Sensor Type: Catalytic sensor ● Detect gas: CH4/Natural gas/H2/ethyl alcohol ● Measure range: 0-100%lel or 0-10000ppm ● Alarm point: 25%lel or 2000ppm ,adjustable ● Accuracy: ≤5%F.S ● Alarm: Voice + vibration ● Language: Support English& Chinese menu switch ● Display: LCD digital display, Shell Material: ABS ● Working voltage: 3.7V ● Battery capacity: 2500mAh Lithium battery ●...

    • Meteorological anemometer wind speed sensor

      Meteorological anemometer wind speed sensor

      Technique Parameter Measurement range 0~45m/s  0~70m/s Accuracy ±(0.3+0.03V)m/s (V: wind speed) Resolution 0.1m/s Staring wind speed ≤0.5m/s Power supply mode DC 5V  DC 12V  DC 24V  Other Out-put Current: 4~20mA  Voltage: 0~2.5V  Pulse:Pulse signal Voltage: 0~5V     RS232 RS485 TTL Level: (frequency; Pulse width) Other Instrument Line length Standard: 2.5m ...

    • Composite portable gas detector

      Composite portable gas detector

      System Description System configuration 1. Table1 Material List of Composite portable gas detector Composite portable Gas Detector USB Charger Certification Instruction Please check materials immediately after unpacking. The Standard is necessary accessories. The Optional is can be choose according to your needs. If you have no need to calibration, set the alarm parameters, or read the...