• LFHC-TBQ Total Radiation Sensor

LFHC-TBQ Total Radiation Sensor

Short Description:

Model:LFHC-TBQ ;

◆Principle:Total radiation sensor uses pyroelectric sensor principle;

Advantages: Precise measurement, fast response, long service life;

Customized :Support;

Payment Methods:Support T/T, L/C, D/P, D/A, O/A, Western Union, Paypal etc.


Product Detail

Product Tags

Application

This sensor is used to measure the spectral range of 0.3-3μm, solar radiation, can also be used to measure the incident solar radiation to the slant of the reflected radiation can be measured, such as the induction downwardly facing, light shielding ring measurable scattered radiation. Therefore, it can be widely applied to the use of solar energy, meteorology, agriculture, building materials, aging and atmospheric pollution and other departments to do the measurement of solar radiation energy.

Technique Parameter

Sensitivity 7~14μV/w.m-2
Spectral range 0.3-3μm
Measuring range 0 ~ 2000W/m2
Power supply DC 5V 
DC 12V 
DC 24V 
Other
Output in the form Pulse: Pulse signal
Current: 4~20mA  
Voltage: 0~2.5V 
Voltage: 1~5V    
Voltage: 0 ~ 20mV
RS232
RS485
Other
Instrument Line length Standard: 2.5m
Other  
Response time ≤ 35 seconds (99%)
Internal resistance approximately 350Ω
Stability ≤ ± 2%
Cosine response ≤7% (solar elevation angle of 10 °)
Azimuth response error ≤5% (solar elevation angle of 10 °)
Temperature characteristics ±2% (-10 ℃ to +40 ℃)
Working ambient temperature -40 ° C ~ +50 ° C
Non-linear ≤ 2%
Weight 2.5kg

Overall structure

LF-0010 TBQ Total Radiation Sensor5
LF-0010 TBQ Total Radiation Sensor4

Transmitter Dimensions

LF-0010 TBQ Total Radiation Sensor6

Installation and use

The sensor should be installed in the space, around the place above the sensing surface without any obstacles. Then the radiation table cable plug is on the north, adjust level position, firmly fixed, then the total radiation sensor output cable acquisition equipment connected to observation. The best cable is securely fixed to the mounting bracket to reduce the fracture or interruptions occur intermittently windy days.

Wiring instructions

Sensor name Wiring color Output Description Corresponding plug pins
TBQ Total radiation sensor Red line Positive signal output Foot 1 of the four-pin plug
Black line Negative signal output Foot 2 of the four-pin plug

RS485 (with address) communication protocol   

1. Serial format
8 data bits
1 stop bit
Parity None
Baud rate 9600 two communication interval of at least 1000ms
2. The communication format
[1] Is written to the device address
Send: 00 10 00 AA (16 hexadecimal data)
Description: 00 - Broadcast address (must be 0); 10 - Write operation (fixed); 00 - Address command (fixed); AA -Write the new address (only,1-255)
Returns: OK (OK return success)
[2] To read the device address
Sent: 000,300 (hexadecimal data)
Description: 00 - Broadcast address (must be 0); 03 - Read operation (fixed);  00 - Address command (fixed)
Returns: Address = XXX (ASCII code data, such as Address = 001, Address = 123, etc.)
Description: Address - address instructions; XXX - Address data, less than three integers, preceded by 0;
[1] Which units followed with a carriage return wrap data, two-byte hexadecimal data 0x0D 0x0A;
[2] The above description ignores transition spaces and '=' character.
[3] Read real-time data
Send: AA 03 0F (16 decimal data)
Description: AA - Device address (only 1-255); 03 - Read operation (fixed); 0F - Data address (fixed)
Returns: The TBQ = XXXX W/m2 (ASCII code data, such as TBQ = 0400 W/m2 TBQ = 1000 W/m2 etc.)
Description: TBQ - TBQ total convergence of f our integers, preceded by 0; W/m2 - units
[1] Which units followed with a carriage return wrap data, two-byte hexadecimal data 0x0D 0x0A.
[2] The above description ignores transition spaces and '=' character.

Instructions for use

The sensor wires in the wiring method according to the instructions in the, and then placed in the position for the measurement of radiation, turn on the power and collection instrument switch, you can get the radiation value of the measurement points.

Precautions

1. Please check the packaging is intact, and check whether the product model is consistent with the selection.
2. Do not live wiring, the wiring is complete check correct before electricity.
3. Do not tamper with the product factory welded components or wires.
4. The sensor is a precision device, use Do not disassemble, contact with the sensor surface with sharp objects or corrosive liquid, so as to avoid damage to the product.
5. Save the certificate and certificate maintenance together with product return.

Troubleshooting

1. When analog output, showing values ​​significantly larger / smaller. Check the pass dirt or debris on the sensor sensing port; if so, with a dry net cloth to wipe away.
2. When analog output the display instrument represents a value of 0 or is not in range to within. May be due to wiring problems lead to acquisition instrument is unable to obtain the correct letter, please check the wiring is correct, firmly.
3. If the above reasons, please contact the manufacturer.

Product maintenance

1. It is not allowed to demolition or loose filter mask, so as not to affect the measurement accuracy. Open or covered with a metal cover especially careful, because the filter mask precious and fragile. Filter mask to maintain smooth, often with a soft cloth or fur smear.
2. Filter mask should not has water, nor should be water vapor condensation hood. Desiccant desiccator should always check whether to change the tide (from blue to red or white), otherwise it must be promptly replaced or desiccant will get the oven to dry so that it changed back to blue again.
3. TBQ total radiation sensor waterproof performance is good, generally a short period of time or smaller precipitation cannot stamped. But the rainfall heavy rains (snow, ice, etc.) or a longer time, for the protection of the radiation table, observers is necessary, according to the specific circumstances of the best stamped, the rain stopped as soon as the cover is open.
4. TBQ total radiation sensor use for more than two years, re-calibration of its sensitivity by the manufacturer or metering department.

Service commitment

TBQ total radiation sensor from the factory on the date one year due to the non-human factors causing quality problems, the production unit is responsible for free repair or replacement. If the user man-made damage, will be charged the cost of fees, but does not charge maintenance fees. In addition, I solemnly promised to be responsible for products manufactured lifelong maintenance.

Selection Table

No.

Power supply

Export

Signal

Explain

LF-0010 -

 

 

TBQ total radiation sensor

(Transmitter)

 

5V-

 

5VPower supply

12V-

 

12VPower supply

24V-

 

24VPower supply

Z-

 

No power

 

V

0-5V

A1

4-20mA

A2

0-20mA

W1-

RS232

W2-

RS485

V3-

0-20mV

X-

Other

E.g.: LF-0010-12V-A: TBQ total radiation sensor 12V power supply, 4-20mA current signal output


  • Previous:
  • Next:

  • Write your message here and send it to us

    Related products

    • Multifunctional Automatic Weather Station

      Multifunctional Automatic Weather Station

      System Components Technical Parameter Working environment: -40℃~+70℃; Main functions: Provide 10-minute instantaneous value, hourly instantaneous value, daily report, monthly report, annual report; users can customize the data collection time period; Power supply mode: mains or 1...

    • Ultrasonic Sludge Interface Meter

      Ultrasonic Sludge Interface Meter

      Features  ● Continuous measurement, low maintenance ● Ultrasonic high frequency technology, stable and reliable performance ● Chinese and English operation interface, easy to operate ● 4~20mA, relay and other interface outputs, system integrated control ● Automatically adjust the transmit power according to the mud layer ● Advanced digital model operation, anti-interference design   ...

    • WGZ-500B, 2B, 3B, 4000B Portable Turbidity Meter

      WGZ-500B, 2B, 3B, 4000B Portable Turbidity Meter

      Features  ● Portable, AC and DC power supply, with low voltage indication and automatic shutdown function. The serial RS232 communication interface can be connected with a micro printer. ● Microcomputer low-power configuration, touch keyboard, LCD screen with backlight, can display date, time, measurement value and measurement unit at the same time. ● Measuring range can be manually selected or automa...

    • CLEAN DO30 Dissolved Oxygen Meter

      CLEAN DO30 Dissolved Oxygen Meter

      Features  ●Boat-shaped floating design, IP67 waterproof grade. ●Easy operation with 4 keys, comfortable to hold, accurate value measurement with one hand. ●Selectable dissolved oxygen unit: concentration ppm or saturation %. ●Automatic temperature compensation, automatic compensation after salinity/atmospheric pressure input. ●User-replaceable electrode and membrane head kit (CS49303H1L) ●Can carry ou...

    • Rain sensor stainless steel outdoor hydrological station

      Rain sensor stainless steel outdoor hydrologica...

      Technique Parameter Water-carrying caliber Ф200 ± 0.6mm Measuring range ≤4mm / min (precipitation intensity) Resolution 0.2mm (6.28ml) Accuracy ± 4% (indoor static test, rain intensity is 2mm / min) Power supply mode DC 5V  DC 12V  DC 24V  Other Output form Current 4 ~ 20mA Switching signal: On-off of reed switch Voltage: 0~2.5V  Voltage: 0~5V     Voltage 1 ~ 5V Other ...

    • Ambient Dust Monitoring System

      Ambient Dust Monitoring System

      System Composition The system consists of particle monitoring system, noise monitoring system, meteorological monitoring system, video monitoring system, wireless transmission system, power supply system, background data processing system and cloud information monitoring and management platform. The monitoring sub-station integrates various functions such as atmospheric PM2.5, PM10 monitoring, ambient...