• Portable compound gas detector

Portable compound gas detector

Short Description:

Thanks for using our portable compound gas detector. Reading this manual will make you quickly master the function and usage of this product. Please read this manual carefully before operating.

Num: Number

Para: Parameter

Cal: Calibration

ALA1: Alarm1

ALA2: Alarm2


Product Detail

Product Tags

System instruction

System configuration

No.

Name

Marks

1

 portable compound gas detector

 

2

Charger

 

3

Qualification

 

4

User manual

 

Please check whether the accessories are complete immediately after received the product. Standard configuration is a must-have for purchasing equipment. Optional configuration is separately configured according to your needs, if you do not need a computer for calibration, setting alarm point, export alarm records. It is not necessary to purchase optional accessories.
System Parameters
Charging time: 3-6 hours
Charging voltage: DC5V
Using time: about 12hours except for alarm status
Detect gas: O2, Combustible gas, CO, H2S, Other gases based on customer's requests
Working environment: Temperature: -20℃ -50℃, Relative humidity: <95%RH(No condensation)
Response time:≤30s(O2); ≤40s(CO); ≤20s(EX);≤30s (H2S)
Size:141*75*43(mm)
Measure range as table 1

Detected gas

Measure Range

Resolution

Alarm Point

Ex

0-100%lel

1%LEL

25%LEL

O2

0-30%vol

0.1%vol

18%vol, 23%vol

H2S

0-200ppm

1ppm

5ppm

CO

0-1000ppm

1ppm

50ppm

CO2

0-5%vol

0.01%vol

0.20%vol

NO

0-250ppm

1ppm

10ppm

NO2

0-20ppm

1ppm

5ppm

SO2

0-100ppm

1ppm

1ppm

CL2

0-20ppm

1ppm

2ppm

H2

0-1000ppm

1ppm

35ppm

NH3

0-200ppm

1ppm

35ppm

PH3

0-20ppm

1ppm

5ppm

HCL

0-20ppm

1ppm

2ppm

O3

0-50ppm

1ppm

2ppm

CH2O

0-100ppm

1ppm

5ppm

HF

0-10ppm

1ppm

5ppm

VOC

0-100ppm

1ppm

10ppm

ETO

0-100ppm

1ppm

10ppm

C6H6

0-100ppm

1ppm

5ppm

Note: The table is for reference only; the actual measurement range is subject to the actual display of the instrument.
Product Characteristics
★ Chinese or English display
★ The compound gas is composed of different sensors, can be flexibly configured to detect up to 6 gases at the same time, and support CO2 and VOC sensors.
★ Three press buttons, sample operation, small size and easy to carry
★ With real time clock, can be set
★ LCD display real time gas concentration and alarm status
★ Large lithium battery capacity, can keep using for long time continuously
★ 3 Alarm type:Audible, vibration, visual alarm, the alarm can be manually muffled
★ Simple automatic zero calibration (just turn on in a non-toxic gas environment)
★ Sturdy and high-grade crocodile clip, easy to carry during operation
★ The shell is made of high-strength special engineering plastic, which is durable, beautiful, and feels good
★ With data storage function, can store 3,000 records, you can view the records on the instrument, or you can connect the computer to export data (optional).

Function introduction

The detector can simultaneously display six kinds of gases numerical indicators. When the gas concentration up to the alarm range, the instrument will automatically conduct alarm action, flashing lights, vibration and sound.
This detector has 3 buttons, one LCD screen, and related alarm system (alarm light, buzzer and shock). It has Micro USB interface that can charge .It also can plug in the USB to TTL adapter to connect to the host computer to calibrate, set alarm parameters or read alarm records.
The instrument itself has a real-time storage function, which can record the alarm status and time in real time. For specific operation instructions and function descriptions, please refer to the description below.
2.1 Buttons function instruction
The instrument has two buttons, function as shown in table 3:
Table 3 Button Function

Marks

Function

Note

 marks1 View parameters,

Enter the selected function

Right button

marks2 Boot, shutdown, please press the button above 3S

Enter the menu and confirm the set value, at the same

Middle button

marks3 Silence

The menu selection button, press the button to enter

Left button

Display
It will go to the boot display by long press the middle keymarks2 in the case of normal gas indicators, shown in Figure1:

Figure 1 Boot display

Figure 1 Boot display

This interface is to wait for the instrument parameters stable. The scroll bar indicates
the waiting time, about 50s. X% is the current progress. The bottom right corner display the real time and power capacity.
When the percentage turns into 100%, the instrument enters the monitor 6 gas display Figure 2:

Figure 2. Monitor 6 gas display interface

Figure 2. Monitor 6 gas display interface

If the user purchases a non-six-in-one, the display interface is different. When three-in-one, there is a gas display position that is not turned on, and the two-in-one only displays two gases.
If you need display one gas interface, you can press the right button to switch. Let's briefly introduce these two gases display interfaces.
1)Multi-gas display interface:
Display: gas type, gas concentration value, unit, status. as shown in Figure 2.

When the gas exceeds the index , the unit's alarm type will be displayed beside the unit (carbon monoxide, hydrogen sulfide, flammable gas alarm type is the first or second level, and the oxygen alarm type is the upper or lower limit), the backlight is on, and the LED light flashes, the buzzer sounds with vibration, and the horn icon vwill appear, as shown in Figure 3.

the interface when alarming

Figure 3. the interface when alarming

Press the left button and clear the alarm sound, the icon change to indicate the alarm status.
2)One gas display interface: 
On the multi-gas detection interface, press the right button and turn to display the gas location interface.

Figure 4 Gas location display

Figure 4 Gas location display

Note: When the instrument is non-six in one, some serial numbers will show [not open]
Press left button and enter one gas display interface.
Display: Gas type, alarm status, time, 1st level alarm value(The lower limit alarm value),2nd level alarm value (The higher limit alarm value), measure range, real time gas concentration, unit.
Below the current gas concentration, it's 'next', press the left buttons turn to next gas’s index, press the left button and switch four kinds gas index. Figure 5, 6, 7, 8 is the four gas parameters. Press back (right button) means switch to detect a variety of gas display interface.

Single gas alarm display shows in Figure 9 and 10

Figure 5 O2

Figure 5 O2  

Figure 6 Combustible gas

Figure 6 Combustible gas

Figure 7 CO

Figure 7 CO   

Figure 8 H2S

Figure 8 H2S

Figure 9 Alarm status of O2

Figure 9 Alarm status of O2 

Figure 10 Alarm status of H2S

Figure 10 Alarm status of H2S

When one gas start alarm, 'next' change to mute. Press the left button and stop alarming, then mute turn to 'next'

Menu Description
When you need set up the parameters, press middle button to enter the menu, main menu interface as Figure 11.

Figure 11 Main menu

Figure 11 Main menu

Icon means selected function, Press left button to select others, Press the right button to enter the function.
Function description:
● Set time: set the time.
● Shut down: close the instrument
● Alarm store: View the alarm record
● Set alarmdata: Set the alarm value, low alarm value and high alarm value
● Calibration: Zero correction and calibration equipment
● Back: back to detect four kinds of gases display.

Set Time
Press left button to select time setting, press right button to enter time setting interface as Figure 12.

Figure 12 Time setting

Figure 13 Year setting

Figure 13 Year setting

Figure 13 Year setting

Icon means select the time for setting, press right button to Figure 13, then press left button to adjust the data, then press right button confirm the data. Press left button to adjust other time data.
Function Description:
Year: setting range 19 to 29.
Month: setting range 01 to 12.
Day: setting range is from 01 to 31.
Hour: setting range 00 to 23.
Minute: setting range 00 to 59.
Back to: Return to the main menu
Shut down
In the main menu, press the left button to select the 'off' function, and then press the right button to shut down. Or long press the right button for 3 seconds
Alarm store
In the main menu, press left button to select 'record' function, then press right button to enter the recording menu, as shown in Figure 14.
● Save Num: total number of storage equipment storage alarm record.
● Fold Num: If the amount of data stored in the device is greater than the total number of storage, it will be overwritten starting from the first data, this item represents the number of overwrites
● Now Num: current data storage number, shown has been saved to No. 326.

Show the latest record first, press the left key to see the next record, and press the right button to return to the main menu, as shown in Figure 14.

Figure 14 Alarm Record Interface

Figure 14 Alarm Record Interface

Figure 15 Specific record query

Figure 15 Specific record query

Show the latest record first, press the left key to see the next record, and press the right button to return to the main menu, as shown in Figure 14.

Alarm Setting
In the main menu interface, press the left button to select the function item of 'alarm setting', and then press the right button to enter the alarm setting gas selection interface, as shown in Figure 16.Press the left key to select the gas type, and press the right button to enter the selected gas alarm value interface. Let’s take carbon monoxide.

Figure 16 Gas Selection Interface

Figure 16 Gas Selection Interface

Figure 17 Alarm Value Setting

Figure 17 Alarm Value Setting

In Figure 17 interface, press the left key choose carbon monoxide "first level" alarm value, then press right key to enter Settings menu, as shown in Figure 18, At this point, press the left button to switch the data bit, press the right button to add the flashing bit value. Set the required value by the left and right keys, and press the middle key to enter the alarm value confirmation interface after setting. At this time, press the left key to confirm. After setting successfully, the position in the bottom middle of the screen shows "setting successfully"; otherwise, it prompts "setting failure", as shown in Figure 19.

Figure 18 Alarm Value Confirmation interface

Figure 18 Alarm Value Confirmation interface

Figure 19 Setting successfully interface

Figure 19 Setting successfully interface

Note: the alarm value set must be less than the factory value (oxygen lower limit must be above than the factory value), otherwise the setting will fail.

Equipment calibration
Note:
1. After the equipment is started, zero correction can be made after initialization.
2. Oxygen in the standard atmospheric pressure can enter the "gas calibration" menu correct display value is 20.9%vol, must not operate "zero correction" in the air.
3. Please do not calibrate the equipment without standard gas.

Zero correction
Step 1: in the main menu interface, press the left button to select the function item of 'device calibration', and then press the right button to enter the calibration password menu, as shown in Figure 20.According to the icon in the last line of the interface, press the left button to switch data bits, press the right button to add 1, enter the password 111111 through the cooperation of the two keys, and press the middle button to switch the interface to the calibration selection interface, as shown in Figure 21.

Figure 20 Password Interface

Figure 20 Password Interface  

Figure 21 Calibration Selection

Figure 21 Calibration Selection

Step 2: press the left key to select items' zero correction function, and then press right key to enter zero calibration menu, through press left key to choose the type of gas to reset, as shown in Figure 22.then press right key to select gas reset menu, confirm the current gas is 0 PPM, press left key to confirm. After successful calibration, 'calibration success' will be displayed in the bottom middle of the screen, while 'failure' will be displayed, as shown in Figure 23.

Figure 22 Gas Selection

Figure 22 Gas Selection

Figure 23 calibration interface

Figure 23 calibration interface

Step 3:Press right key to return to the gas type selection interface after the completion of zeroing correction. At this time, other gas types can be selected for zeroing correction. The method is the same as above. After zero, return to the detection gas interface step by step or wait 15 seconds, the instrument will automatically return to the detection gas interface.

Full calibration
Step1:After the gas to be stable display value, enter the main menu, call up the Calibration menu selection. The specific methods of operation like the step one of cleared calibration.
Step 2: Select 'gas calibration' feature items, press the right key to enter Calibration value interface, then set the concentration of standard gas through the left and right key, suppose now that Calibration is carbon monoxide gas, the concentration of Calibration gas concentration is 500ppm,at this time set to '0500 'can be. As shown in Figure 25.

Figure 24  Gas Selection

Figure 24  Gas Selection

Figure 25 Set the value of standard gas

Figure 25 Set the value of standard gas

Step 3: After setting the calibration,holding down the left button and right button, change the interface to the gas calibration interface, as shown in Figure 26,this interface has a current value detected gas concentration. When the countdown goes to 10, you can press left button to manual calibration, after the 10S, the gas automatic calibrates, after the Calibration is successful, the interface displays' Calibration success! 'On the contrary Show' Calibration Failed! '.The display format shown in Figure 27.

Figure 26 Calibration Interface

Figure 26 Calibration Interface

Figure 27 Calibration results

Figure 27 Calibration results

Step4: After Calibration is successful, the value of the gas if the display is not stable, You can select 'rescaled', if the calibration fails, check the calibration gas concentration and calibration settings are the same or not. After calibration of the gas is complete, press the right to return to the gas detection interface.

Step 5: after all the gas calibration is completed, press the menu to return to the detection gas interface, or automatically to return to the gas detection interface.

Back
In the main menu interface, press the left key to select the 'back' function item, and then press the right button to return to the previous menu

Note

1) Be sure to avoid a long-time charging. The charging time may extends, and the instrument's sensor may be affected by differences in the charger (or charging environmental differences) when the instrument is open. In most severe cases, it may even appear instrument error display or alarm situation.
2) The normal charging time of 3 to 6 hours or so, try not to charge the instrument in six hours or more to protect the effective life of the battery.
3) The instrument can work for 12 hours or so after fully charged (except for the alarm state, because the flash when the alarm, vibration, sound require additional power. The working hours reduced to 1/2 to 1 / 3 when keeping alarm status).
4) When the power of the instrument is very low, the instrument will be turned on and shut down automatically frequently. At this time, it is necessary to charge the instrument
5) Be sure to avoid using the instrument in a corrosive environment
6) Be sure to avoid contact with water instrument.
7) It's should be unplug the power cable, and charged every 2-3 months, in order to protect the normal battery life when unused for long time.
8) If the instrument crash or cannot be opened, you can unplug the power cord, then plug the power cord to relieve accident crash situation.
9) Make sure the gas indicators are normal when open the instrument.
10) If you need to read the alarm record, its best enter the menu to accurate time before the initialization has not completed to prevent confusion when reading records.


  • Previous:
  • Next:

  • Write your message here and send it to us

    Related products

    • Single Gas Detector User’s

      Single Gas Detector User’s

      Prompt For security reasons, the device only by suitably qualified personnel operation and maintenance. Prior to the operation or maintenance, please read and fully manage all solutions to these instructions. Including operations, maintenance of equipment and process methods. And a very important safety precautions. Read the following Cautions before using the detector. Table 1 Cautions Cautions ...

    • Single-point Wall-mounted Gas Alarm

      Single-point Wall-mounted Gas Alarm

      Structure chart Technical parameter ● Sensor: electrochemistry, catalytic combustion, infrared, PID...... ● Responding time: ≤30s ● Display mode: High brightness red digital tube ● Alarming mode: Audible alarm -- above 90dB(10cm) Light alarm --Φ10 red light-emitting diodes (leds) ...

    • Portable pump suction single gas detector

      Portable pump suction single gas detector

      System Description System configuration 1. Table1 Material List of Portable pump suction single gas detector Gas Detector USB Charger Please check materials immediately after unpacking. The Standard is necessary accessories. The Optional can be chosen according to your needs. If you have no need to calibrate, set the alarm parameters, or read the alarm record, don’t purchase the optional acc...

    • Composite portable gas detector

      Composite portable gas detector

      System Description System configuration 1. Table1 Material List of Composite portable gas detector Composite portable Gas Detector USB Charger Certification Instruction Please check materials immediately after unpacking. The Standard is necessary accessories. The Optional is can be choose according to your needs. If you have no need to calibration, set the alarm parameters, or read the...

    • Digital gas transmitter

      Digital gas transmitter

      Technical Parameters 1. Detection principle: This system through standard DC 24V power supply, real-time display and output standard 4-20mA current signal, analysis and processing to complete digital display and alarm operation. 2. Applicable objects: This system supports the standard sensor input signals. Table 1 is our gas parameters setting table (For reference only, users can set the parameters a...

    • Compound single point wall mounted gas alarm

      Compound single point wall mounted gas alarm

      Product Parameters ● Sensor: Combustible gas is catalytic type, other gases are electrochemical, except special ● Responding time: EX≤15s; O2≤15s; CO≤15s; H2S≤25s ● Work pattern: continuous operation ● Display: LCD display ● Screen Resolition:128*64 ● Alarming mode: Audible & Light Light alarm -- High intensity strobes Audible alarm -- above 90dB ● Output control: relay output with two wa...